doi: 10.1007/bf02483698
pmid: 4729093
A constant-current, 1 kHz electrical-shock stimulator is described. It can be battery-powered or run from a well isolated power supply (also described). The device can supply up to 25 mA to loads from zero to 16 kω, with reduced output at higher resistive loads. A tenfold change in load produces less than a 1 mA change in output current. In the past year, several devices of this design have been built and have functioned extremely reliable (Chapman andSmith, 1972), with several types of commercially available skin electrodes.
Electroshock, Methods, Humans, Electric Stimulation, Electronics, Medical
Electroshock, Methods, Humans, Electric Stimulation, Electronics, Medical
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 1 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |