doi: 10.26464/epp2023004
Ground calibration experiments of the Mars orbiter magnetometer (MOMAG) onboard the orbiter of Tianwen-1 were performed to determine the sensitivity, misalignment angle, and offset of the sensors. The linearity of the applied calibrated magnetic fields and the output from the sensors were confirmed to be better than 10−4, and the sensor axes were orthogonal to each other within 0.5 degrees. The temperature dependencies of the sensitivity and misalignment angle were examined, but no clear signatures of temperature dependencies could be seen. Sensor offset and the stability of sensor offset drift with a temperature change were also determined by the rotation method. The stability of the sensor offset drift was less than 0.01 nT/°C. The ground calibration of MOMAG determines all the calibration parameters of the sensors for accurate magnetic field measurements in orbit with the appropriate corrections.
Environmental sciences, QC801-809, ground calibration, Science, tianwen-1, Q, Geophysics. Cosmic physics, GE1-350, mars orbiter magnetometer
Environmental sciences, QC801-809, ground calibration, Science, tianwen-1, Q, Geophysics. Cosmic physics, GE1-350, mars orbiter magnetometer
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 1 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |