Spectral Doppler interrogation of flow across a patent foramen ovale (PFO) allows recording of the instantaneous pressure gradient between left and right atrium (RA). The assessment of RA pressure using the size and collapsibility of the inferior vena cava would thus allow estimation of left atrial (LA) pressure. In this article, we illustrate the value of spectral Doppler interrogation of flow across the PFO by transthoracic echocardiography as a novel and simple tool for the assessment of LA pressure and left cardiac hemodynamics in addition to the conventional noninvasive parameters.
Adult, Adolescent, Hemodynamics, Foramen Ovale, Patent, Humans, Middle Aged, Echocardiography, Doppler, Color
Adult, Adolescent, Hemodynamics, Foramen Ovale, Patent, Humans, Middle Aged, Echocardiography, Doppler, Color
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 3 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |