The digitsal voltmeter (DVM) input and leakage resistances have been analysed in the DVM-based resistance measurement system using one or two DVMs. The results are given both theoretically and experimentally. Shunting errors due to DVM-input and leakage resistances were substantially reduced by using a proper measurment method. The method is being further investigated for measuring high-ohm standard resistors
resistance measurement, Input resistance, Digital Voltmeters; Input resistance; leakage resistance; resistance measurement, leakage resistance, Digital Voltmeters
resistance measurement, Input resistance, Digital Voltmeters; Input resistance; leakage resistance; resistance measurement, leakage resistance, Digital Voltmeters
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 7 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |