Vibration in the audio frequency band affects the performance of rotating gravity gradiometers used for airborne mineral exploration. This is probably due to translation to rotation coupling inside the gradiometer platform. It was found that the DC gravity gradient signal was proportional to the square of the third time derivative of position, or jerk squared. The demanding airborne environment for such instrumentation demands a light weight broadband acoustic shield and vibration isolator. This paper presents the design principles for such an isolator, based on vibration isolated spherical shell structures. Performance data are presented as well as flight test data that demonstrated a 14% gravity gradient noise reduction compared with an unshielded instrument.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 6 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |