doi: 10.1002/rcm.4400
pmid: 20049885
Abstract The on‐line detection of gaseous peroxyacetyl nitrate (PAN) using selected ion flow tube mass spectrometry (SIFT‐MS) has been investigated using a synthetic sample of PAN in air at a humidity of ∼30%. Using the H 3 O + reagent ion, signals due to PAN at m/z 122, 77 and 95 have been identified. These correspond to protonated PAN, protonated peractetic acid and its water cluster, respectively. These products and their energetics have been probed through quantum mechanical calculations. The rate coefficient of H 3 O + has been estimated to be 4.5 × 10 −9 cm 3 s −1 , leading to a PAN sensitivity of 138 cps/ppbv. This gives a limit of detection of 20 pptv in 10 s using the [M+H] + ion of PAN at m/z 122. Copyright © 2010 John Wiley & Sons, Ltd.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 15 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |