Dataset containing drift of selected Pt-Rh thermocouples which are part of a multi-wire thermocouple. Drift is in emf terms and in temperature terms. Drift was measured by exposing the thermocouples to a temperature of approximately 1324 °C, with periodic in-situ calibration using a Co-C 'fixed-point' point cell (melting point 1324.29 °C).
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |