The purpose of this work was to evaluate computed tomography simulators used in radio-therapy treatment planning in Serbia, Croatia, and Bosnia and Herzegovina. A survey of quality assurance programmes of 24 computed tomography simulators in 16 facilities was conducted. A dedicated CT-to-ED phantom was scanned at 120 kV and 140 kV, to obtain CT-to-ED conversion curves as well as CTDIvol. Thoracal phantoms were scanned in the standard and extended field of view to evaluate the dosimetric effect on treatment planning and delivery. The mean age of the measured scanners was 5.5 years. The mean water HU value was ?6.5 (all scanners, all voltages) and air HU value was ?997. Extended field of view computed tomography data differ from the standard field of view and differences between conversion curves have significant dosimetric impact. The CTDI data showed a large range of values between centers. Better quality assurance of computed tomography simulators in all countries is recommended. The CT-to-ED curve could be used as default at one voltage and per manufacturer. Extended field of view imaging can be used, but treatment planning should be avoided in the regions out of the standard field of view.
radiotherapy tratment planning, quality assurance, QC770-798, conversion curve, conversion curve CT-to-ED, radiotherapy treatment planning, CT simulator, CT-to-ED, Nuclear and particle physics. Atomic energy. Radioactivity, computed tomography simulator, QC, radiotherapy, CT simulator, QC, CT-to-ED, conversion curve ct-to-ed
radiotherapy tratment planning, quality assurance, QC770-798, conversion curve, conversion curve CT-to-ED, radiotherapy treatment planning, CT simulator, CT-to-ED, Nuclear and particle physics. Atomic energy. Radioactivity, computed tomography simulator, QC, radiotherapy, CT simulator, QC, CT-to-ED, conversion curve ct-to-ed
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 1 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |
| views | 67 | |
| downloads | 75 |

Views provided by UsageCounts
Downloads provided by UsageCounts