doi: 10.1117/12.3046216
We demonstrate a 3.2 MHz-OCT system for inter-volumetric dynamic optical coherence tomography of ex vivo porcine kidney tissue. Employing a home-built Fourier Domain mode locking (FDML) laser with a 1310 nm wavelength, the system achieved a lateral resolution of 3.48 μm and a frame rate of 612 Hz. A motorized XYZ positioning stage enabled the precise acquisition of multiple volumes, which were seamlessly stitched together to generate a comprehensive dataset with a total area of 2.6 × 2.6 mm2. Validations against histological sections confirmed the system’s ability to visualize cellular tissue structures.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 3 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |