Bone Strain Index (BSI) is a new finite element analysis tool applied to hip dual energy X-ray absorptiometry scans. The aim of this study was to assess the short-term precision error of BSI on the proximal femur, both on a phantom and patients. The International Society for Clinical Densitometry guidelines were followed for short-term precision error assessment. Dual energy X-ray absorptiometry measurements were performed on an anthropomorphic femur phantom that was scanned twice for 30 times, for a total of 60 scans. For the in vivo part, 30 subjects were scanned twice. BSI precision error was compared to that of bone mineral density (BMD). Both for the phantom and the in vivo study BSI reproducibility was lower compared to that of BMD, as the precision error of BSI resulted 3 times higher compared to that BMD. For phantom measurements, the highest precision value was that of total femur (TF) BMD (coefficient of variation [CoV] = 0.63%, reproducibility = 98.24%), while the lowest precision was the femoral neck (FN) BSI (CoV = 3.08%, reproducibility = 91.48%). Similarly, for the in vivo study, the highest precision was found at TF BMD (CoV = 1.36%, reproducibility = 96.22%), while the lowest value of precision was found for FN BSI (CoV = 4.17%, reproducibility = 88.46%). Reproducibility at TF was always better compared to that of the FN. BSI precision error was about 3 times higher compared to BMD, confirming previous results of lumbar spine BSI. The main source of variability of this new software is related to patient positioning.
Bone strain index; dual energy X-ray absorptiometry; finite element analysis; finite element method; osteoporosis, Male, Phantoms, Imaging, Finite Element Analysis, Reproducibility of Results, Patient Positioning, Absorptiometry, Photon, Bone Density, Humans, Osteoporosis, Female, Femur, Software, Aged
Bone strain index; dual energy X-ray absorptiometry; finite element analysis; finite element method; osteoporosis, Male, Phantoms, Imaging, Finite Element Analysis, Reproducibility of Results, Patient Positioning, Absorptiometry, Photon, Bone Density, Humans, Osteoporosis, Female, Femur, Software, Aged
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 13 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |
| views | 3 |

Views provided by UsageCounts