AbstractPurposeTo implement and validate a beam current transformer as a passive monitoring device on a pulsed electron beam medical linear accelerator (LINAC) for ultra‐high dose rate (UHDR) irradiations in the operational range of at least 3 Gy to improve dosimetric procedures currently in use for FLASH radiotherapy (FLASH‐RT) studies.MethodsTwo beam current transformers (BCTs) were placed at the exit of a medical LINAC capable of UHDR irradiations. The BCTs were validated as monitoring devices by verifying beam parameters consistency between nominal values and measured values, determining the relationship between the charge measured and the absorbed dose, and checking the short‐ and long‐term stability of the charge‐absorbed dose ratio.ResultsThe beam parameters measured by the BCTs coincide with the nominal values. The charge‐dose relationship was found to be linear and independent of pulse width and frequency. Short‐ and long‐term stabilities were measured to be within acceptable limits.ConclusionsThe BCTs were implemented and validated on a pulsed electron beam medical LINAC, thus improving current dosimetric procedures and allowing for a more complete analysis of beam characteristics. BCTs were shown to be a valid method for beam monitoring for UHDR (and therefore FLASH) experiments.
FLASH RT, ultra-high dose rate, Clinical Protocols, Radiation Monitoring, Humans, Electrons, Radiotherapy Dosage, Particle Accelerators, beam current monitoring, Radiometry, Radiation Measurements
FLASH RT, ultra-high dose rate, Clinical Protocols, Radiation Monitoring, Humans, Electrons, Radiotherapy Dosage, Particle Accelerators, beam current monitoring, Radiometry, Radiation Measurements
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 53 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 1% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 1% |
| views | 9 | |
| downloads | 25 |

Views provided by UsageCounts
Downloads provided by UsageCounts