Controller for an axial-flow blood pump.

Authors: D V, Amin; J F, Antaki; P, Litwak; D, Thomas; Z, Wu; Y C, Yu; S, Choi; +2 Authors

Controller for an axial-flow blood pump.

Abstract

An axial-flow ventricular assist device (VAD) under development at the authors' facility is intended for use as a long-term implantable device. At high speeds axial-flow VADs can collapse the native ventricle and damage the heart muscle, lung tissue, and blood. A prototype algorithm was developed to maintain physiologic perfusion to the vital organs while preventing ventricular collapse, through analysis of the electrical current waveform of the motor. The premise of the control algorithm is that the hemodynamics of the patient are reflected in the shape of this waveform. This approach is intended to eliminate the need for invasive sensors, thus effectively using the pump itself as a transducer. The control algorithm regulates the speed of the pump by comparing the motor-current waveform with reference waveforms using a matched filter. The matched filter was evaluated by its classification and differentiation performance. Thus far, the authors have been able to classify the waveforms into one of the four physiologic regions (below, within, or above the optimal range, and ventricular suction) with over 90% reliability. Ongoing work is directed toward improving the detection of ventricular suction, as this condition must be strictly avoided.

Related Organizations
Keywords

Transducers, Hemodynamics, Humans, Signal Processing, Computer-Assisted, Heart-Assist Devices, Prosthesis Design, Algorithms

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    0
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Average
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Average
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Average
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
0
Average
Average
Average
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!