This study evaluated the ability of practitioners experienced in the use of Brãnemark implant components to impart a desired torque using handheld screwdrivers. The torque output and variability of mechanical torque devices was also examined. For clinicians using handheld screwdrivers, values ranged from 0.7 to 18.1 Ncm, 1.4 to 33.7 Ncm, and 8.2 to 36.2 Ncm for the 10-Ncm, 20-Ncm, and 32-Ncm torques respectively. The manually operated mechanical torquing devices produced torque values within the tolerances specified by their respective manufacturers at all levels. Significant variation from the manufacturer's set values were noted as a function of tightening speed with the electronic torquing device. The data indicate that there is wide variation in the ability of clinicians to perceive adequate torquing forces applied to implant components. Calibrated torquing devices are mandatory if proper torquing procedures are to be accomplished.
Dental Implants, Dental Stress Analysis, Calibration, Dental Implantation, Endosseous, Humans, Dental Abutments, Stress, Mechanical
Dental Implants, Dental Stress Analysis, Calibration, Dental Implantation, Endosseous, Humans, Dental Abutments, Stress, Mechanical
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 61 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |