Determination of flow through splenorenal shunts performed for portal hypertension was tried in 7 patients using a dye dilution technique. Varying shunt flows were obtained. The results of the flow determinations were compared with the findings at arteriography and the cause of the varying flows could be revealed. Retrograde flow through the left ovarian vein leading to incomplete mixing was the most common source of error. The employed dye dilution techniques lends itself to shunt flow determination only if sources of error can be ruled out by arteriography.
Radiography, Umbilical Veins, Evaluation Studies as Topic, Portal Vein, Regional Blood Flow, Splenic Vein, Hypertension, Portal, Dye Dilution Technique, Humans, Renal Veins
Radiography, Umbilical Veins, Evaluation Studies as Topic, Portal Vein, Regional Blood Flow, Splenic Vein, Hypertension, Portal, Dye Dilution Technique, Humans, Renal Veins
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 1 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |