[Calculation of water equivalent diameter based on anteroposterior localizer CT images].

Authors: Baohui, Liang; Yajun, Fu; Mingna, Cao;

[Calculation of water equivalent diameter based on anteroposterior localizer CT images].

Abstract

ObjectiveTo explore a method for calculating water equivalent diameter (Dw) based on localizer CT images for calculation of the size specific dose estimates (SSDE).MethodGE Revolution CT and LightSpeed VCT were used to scan CT dose index phantoms 16 cm and 32 cm in diameter at the tube voltages of 80, 100 and 120 kV to obtain the axial image and anteroposterior localizer radiograph. According to the definition of CT Hounsfield unit, the axial images were used to calculate the conversion factors that convert the phantom thickness to water equivalent thickness. The gray value of the localizer radiograph and the water equivalent thickness were calibrated with a linear equation, and the parameters of the calibration were used to calculate the water equivalent thickness. The method was verified using 2 CT dose index phantoms and in 22 patients undergoing chest and abdominal CT examination.ResultComparison of the water equivalent diameter (Dw) based on the localizer radiograph and axial image of the 2 phantoms showed that the percentage difference between Dw from the axial images and from the localizer radiograph was below 3%. The trend of Dw variations with location in the two methods was sonsistent. The difference in Dw in intermediate region of interest between the axial image and the localizer radiograph from the 22 patients was below 6.6%. With the mean Dw in the ROI, the maximum percentage difference was 7.5%.ConclusionCalibration of the gray value of the localizer radiograph and the water equivalent thickness using the axial image and localizer radiograph of CT dose index phantoms allows quick calculation of the SSDE based on the parameters of calibration.

Related Organizations
Keywords

Phantoms, Imaging, Calibration, Humans, Water, Radiation Dosage, Tomography, X-Ray Computed

  • BIP!
    Impact byBIP!
    citations
    The total number of citations received by all articles of the researcher of interest.
    0
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Average
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Average
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Average
Powered by OpenAIRE graph
Found an issue? Give us feedback
citations
The total number of citations received by all articles of the researcher of interest.
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
0
Average
Average
Average
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!