The determination of glucose is possible with the enzymatic reaction of glucose oxidase and potentiometric detection. The signal is proportional to the concentration up to 50 mg/dl. This value is fixed by the concentration of oxygen in the sample. By adding catalase, concentrations up to 2000 mg/dl are detectable. The steepness of the calibration curve is not affected by oxygen concentrations greater than 4 mg/l. In contrast to amperometric sensors, an influence of deposits on the electrodes surface on the signal cannot be found with potentiometric sensors.
Oxygen, Glucose Oxidase, Glucose, Potentiometry, Biosensing Techniques, Hydrogen-Ion Concentration, Catalase, Electrodes