Total joint replacement is used for treating osteoarthritis of the trapeziometacarpal joint. The prosthesis is a ball and socket design resembling a total hip prosthesis and with the same risk of head luxation. To reduce this risk, dual mobility articulations have also been introduced in prosthesis of the thumb, and this is a case report of luxation of the polyethylene liner in a dual mobility prosthesis, illustrating that introducing new designs in prosthesis for the basal joint of the thumb may lead to similar complications as are seen in the hip joint.
Arthroplasty, Replacement, Hip, Joint Prosthesis, Joint Dislocations, Humans, Hip Prosthesis, Prosthesis Design, Prosthesis Failure