Monopolar cautery and adverse effects on cochlear implants.

Authors: Anita, Jeyakumar; Meghan, Wilson; Jonathan E, Sorrel; J Benjamin, McIntire; Dara D, Jones; Todd M, Brickman; Moises, Arriaga;

Monopolar cautery and adverse effects on cochlear implants.

Abstract

The use of monopolar cautery has been widely regarded as a contraindication in the setting of a cochlear implant because of presumed risk to the implant device. There are very limited data to support this contraindication.(1) To evaluate the effects of monopolar cautery on cochlear implant devices. (2) To determine whether monopolar cautery changes the endocochlear temperature in an implanted cochlea.Sixteen cochlear implants from 3 manufacturers (Advanced Bionics LLC, Cochlear Americas Ltd, and Med-El Ltd) were implanted into 2 unembalmed, fresh cadavers. Monopolar cautery was applied to either the tongue or abdomen at coagulation settings of 10 W or 50 W for 30 minutes. Impedance and integrity testing were performed before, during, and after 30 minutes of cautery. The temperature in the endocochlear perilymph was measured during cautery. After explantation, devices were returned to the manufacturer for an in-depth "failure" analysis according to each manufacturer's current protocol evaluating explanted devices.Basic science laboratory, tertiary medical center.Cadaveric study.Application of monopolar cautery to implanted cochlear implants in a cadaveric model.(1) Changes to the implanted devices either during electrocautery or following failure analysis. (2) Changes in the intracochlear temperature.No change in impedance, integrity testing, or failure analysis occurred at any cautery setting when applied to either the oral cavity or abdomen. The temperature of the cochlea did not increase with up to 30 minutes of cautery at a setting of 50 W. Comprehensive device analysis did not show any evidence of device damage at the conclusion of the study.Monopolar cautery did not produce detectible damage to any of the cochlear implant devices or produce detectible temperature change in the cochlea at low or high levels of cautery in the oral cavity in this experimental model.

Related Organizations
Keywords

Equipment Failure Analysis, Cochlear Implants, Tongue, Abdomen, Cautery, Cadaver, Temperature, Humans, Perilymph

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    19
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Top 10%
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Top 10%
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Top 10%
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
19
Top 10%
Top 10%
Top 10%
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!