To establish a UPLC characteristic chromatographic profile analysis method to quickly assess Poria quality and provide basis fro controlling Poria quality. The UPLC characteristic chromatographic profiles of fifteen batches of Poria were determined by ACQUITY UPLC, with HSS T3 Column (2.1 mm x 100 mm, 1.8 microm) eluted with the mobile phases of water containing 0.05% phosphoric acid and acetonitrile in gradient mode. The detection wavelength was set at 243 nm. The common mode of the UPLC characteristic chromatographic profile was set up. There were 20 common peaks, seven of which were identified, and the similar degrees of the fifteen samples to the common mode were between 0.787 and 0.974. The method was so time-saving that it can be used for the quality control of Poria.
Acetonitriles, Phosphoric Acids, Poria, Chromatography, High Pressure Liquid
Acetonitriles, Phosphoric Acids, Poria, Chromatography, High Pressure Liquid
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |