Multi-institutional evaluation of the Tokyo University Ventricular Assist System.

Authors: N, Sato; H, Mohri; Y, Sezai; H, Koyanagi; I, Fujimasa; K, Imachi; K, Atsumi; +2 Authors

Multi-institutional evaluation of the Tokyo University Ventricular Assist System.

Abstract

A total of 61 VASs developed in Tokyo University were evaluated at 21 institutions in the period 1985-1989 for determination of its reliability and effectiveness. The system is comprised of a pneumatic sack-type pump (Nippon Zeon Co.), and its driving console (Aishin Seiki Co.). The stroke volume of the pump is 40 ml and blood contacting surfaces are coated with Cardiothane. Ages of the patients (pts) ranged from 12-82 yrs (mean 58 yrs). VASs were used in the assist mode of LVAD (54 pts), RVAD (5 pts) and BVAD (2 pts). Most of the cases (58 pts) included postocardiotomy cardiogenic shock after surgery for ischemic (28 pts), valvular (22 pts), both ischemic and valvular (7 pts) and congenital (1 pt) heart diseases. Average duration of the assist ranged from 1 hr-20 days (mean 5.6 dys). The VADs could be weaned in 34 cases (56%) and among these, 13 cases (21%) survived to discharge from the hospital. Causes of death in cases which could be weaned from the VAD included multiple organ failure/due to delayed institution of adequate circulatory support, renal failure and systemic infection. Small and minute thrombus formations were noted in 7 cases however, no pump originated thromboembolism were complicated. No troubles of the pump including leakage nor breakage, no mechanical failures of the driving consoles were experienced in any of the cases. Thus, it is concluded that the system was proved to be clinically effective and reliable.

Related Organizations
Keywords

Adult, Aged, 80 and over, Heart Failure, Male, Adolescent, Heart Diseases, Multiple Organ Failure, Shock, Cardiogenic, Thrombosis, Equipment Design, Middle Aged, Postoperative Complications, Humans, Equipment Failure, Female, Heart-Assist Devices, Child, Aged

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    10
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Average
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Top 10%
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Average
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
10
Average
Top 10%
Average
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!