During an animal study using five bastard dogs the hemodynamic relations during extracorporeal perfusion with an arteriovenous circuit and extracorporeal CO2 removal (ECCO2-R) were analyzed. We utilised a prototype circuit (Promed, West-Germany) which is by diminution and using silicon rubber adapted to the application in newborn. A silicon lung (Scimed, 0.8 m2, ECCO2-R) was used in the circuit. During the six hours perfusion period a main arterial blood pressure of 83 +/- 17 mmHg and an extracorporeal blood flow of 412 +/- 56 ml/min was found using arteriovenous perfusion via the femoral artery and vein. We found a statistical significant correlation (r = 0.66) between the main arterial blood pressure and the extracorporeal blood flow. Further more there was a significant correlation between the arterial PCO2 and the blood flow through membrane lung (r = 0.81). In conclusion our animal experiments approved the possibility of hemodynamic adaptation to an arteriovenous circuit as well as effectiveness of CO2 removal via such circuit.
Oxygen, Dogs, Extracorporeal Membrane Oxygenation, Pulmonary Gas Exchange, Silicones, Animals, Membranes, Artificial, Carbon Dioxide, Blood Flow Velocity, Oxygenators, Membrane
Oxygen, Dogs, Extracorporeal Membrane Oxygenation, Pulmonary Gas Exchange, Silicones, Animals, Membranes, Artificial, Carbon Dioxide, Blood Flow Velocity, Oxygenators, Membrane
| citations The total number of citations received by all articles of the researcher of interest. | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |