pmid: 1739190
Total respiratory resistance (Rrs) was measured by the application of a sine wave of airflow to the mouth at an oscillation frequency of 10 Hz. The instrument used was the Siemens Siregnost FD5. The Rrs data were correlated with height, age, sex, and weight in 73 patients and 29 healthy subjects. The patients took part in a rehabilitation program for restoration of their locomotion function. Both groups had normal lung function (VC, FEV1) and no signs of pulmonary disease. The only important determining factor for the value of the Rrs was height. The mean Rrs of 102 subjects was 0.29 +/- 0.08 kPa.1-1.s. Other studies gave values between 0.23 +/- 0.05 and 0.32 +/- 0.10 kPa.1-1.s.
Adult, Male, Airway Resistance, Body Weight, IMPEDANCE, Body Height, Respiratory Function Tests, Reference Values, Oscillometry, Humans, Regression Analysis, Female, OBSTRUCTION, SMOKING, Child, REACTANCE
Adult, Male, Airway Resistance, Body Weight, IMPEDANCE, Body Height, Respiratory Function Tests, Reference Values, Oscillometry, Humans, Regression Analysis, Female, OBSTRUCTION, SMOKING, Child, REACTANCE
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 16 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |