[Avoidance of intermittent T-wave oversensing with device programming].

Authors: M, Wieczorek; R, Hoeltgen; I, Djajadisastra;

[Avoidance of intermittent T-wave oversensing with device programming].

Abstract

We report the case of a 35-year-old man who was suffering from severe heart failure due to cardiomyopathy. He underwent heart transplantation years ago and developed complex ventricular arrhythmias in the following months in combination with recurrent episodes of syncope due to hypertrophic non-obstructive cardiomyopathy in the transplanted heart, so a dual chamber ICD was implanted. Months later repetitive episodes of intermittent T-wave oversensing with consecutive activation of the ICD could be observed. Surgical revision of the electrode was performed and the patient was closely followed up. One year later, further episodes of T-wave oversensing led to multiple inappropriate IDC-shocks. A very short AV-conduction time was programmed to allow ventricular capture whenever possible, because T-wave oversense after ventricular capture would be annotated as single ventricular ectopy not resulting in antitachycardia pacing. As a consequence, the patient was free from inappropriate ICD-shocks, but showed several shorter episodes of T-wave oversensing. They were all initiated by atrial activity that was seen in the refractory period, thus leading to a loss of AV synchrony. Programming a very short post ventricular atrial refractory period (PVARP) in addition to a short AV-delay led to the complete disappearance of T-wave oversensing in this patient. During a 9-month follow-up, no further tachycardia episodes were detected by the device.

Keywords

Adult, Male, Electrocardiography, Therapy, Computer-Assisted, Tachycardia, Ventricular, Humans, Defibrillators, Implantable

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    3
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Average
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Average
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Average
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
3
Average
Average
Average
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!