In this paper, a simple and highly sensitive laser fluorimetric system is described, and the value of time discrimination for suppressing unwanted signals and decreasing the minimum detectable concentration is demonstrated. It is confirmed that time discrimination is very useful for ultratrale determination of fluorescent complex. This technique is successful with the combined use of a pulsed laser and a pulse-gate signal detection system for samarium complex. The linear range and the detectin limit of the method are 0. 01-10microg/ml and 3.3ng/ml, respectively. The RSD for Sm3+ (0. 5microg/ml) ( n =12) is 3.7%. The method was used to detemine samarium in rare earth mixtures. The recovery was in the range of 90.0-110.0%, and the mean recovery was 101.4% ( n = 6 )
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |