1) The implanted TAH offers a method of evaluating the effects of varying physiologic demands cardiac output and venous return. 2) Cardiac output and venous return measurements offer a method of evaluating system design of the device and driving and control-logic improvements in awake, unanesthetized animals. 3) The obtained cardiac function and venous return curves suggest that the present TAH is less responsive to increases in right atrial pressure than the natural heart, but is still completely controlled by venous return. Therefore, it is extremely important to make every effort to eliminate any factor causing excessive venous return.
Animals, Blood Pressure, Cattle, Heart, Heart, Artificial, Cardiac Output, Venous Pressure
Animals, Blood Pressure, Cattle, Heart, Heart, Artificial, Cardiac Output, Venous Pressure
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 8 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |