Our simple liquid chromatographic system for splitless gradient nanocolumn separations at backpressure up to 50 bars was modified to run analysis under the pressure up to 300 bars. Programmable syringe pump was equipped with a 250-L high-pressure syringe and mobile phase gradients were created and stored in relatively long capillary. Available linear gradients of volumes from 5 to 50 L traced by uracil are presented. Sample introduction was performed as separate step and real large volume of sample (tens of microliters) could be injected and preconcentrated on-column. Relative standard deviation of retention times and peak areas were received (RSD < 0.3 % and RSD < 5 % respectively).
liquid chromatography, nano column gradient separations, gradient separation