Breaking of surgical drill bits and subsequent dislodging in the bone are quite common in the field of orthopedics. Even though a few methods have been reported to remove dislodged drill bits, we present a novel method to remove a broken drill bit without additional instruments or a secondary incision. A broken cannulated drill bit within the locking screw hole inside the neck of a femur was retrieved using a depth gauge with a curved tip that hooked onto the edge of the drill bit. By employing a clockwise and counter-clockwise twisting, the broken drill bit was retrieved through the proximal reaming tract. The 4mm tract, which was established by proximal reaming, immensely facilitated safe and time-efficient removal of the drill tip without further trauma or prolonging the surgery time. With our technique, the removal was simple and safe without further soft tissue trauma and blood loss. We advocate this approach for implementation in similar cases.
Orthopedics
Orthopedics
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 2 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |