
Abstract Vibrating mesh atomizers (VMA) are increasing in demand for various aerosol applications due to their ability to generate uniformly sized droplets. Currently there are two types of VMA (commercial metallic membranes and silicon based). High Uniformity and control of small droplet size are the basic requirements for many aerosol applications, for which ultrasonic or VMA are employed. However, there is limited research on understanding the droplet size distribution of different types of atomizers. In this study three aerosol generators were investigated: Ultrasonic, metallic VMA, and MEMS-based silicon VMA. The primary objective was to compare these devices on droplet size distribution and mechanism of action. A systematic study to compare the performance of the two VMA was investigated based on droplet distribution, volumetric median diameter (VMD) using liquids with different physiochemical properties. Size distribution of the droplet produced by the metallic VMA was twice the span compared to silicon VMA for fluids with viscosity <2cP. The metallic VMA also resulted in an increase in VMD as the viscosity increased, whereas the Si VMA did not see a significant increase in VMD. The silicon-based VMA demonstrated a 4-15x increase in fine particle fraction control compared to metallic VMA. The results demonstrate that silicon based VMA has narrower droplet distribution with more uniform droplet size and lower span compared to metallic VMA.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 24 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |
