abstractWe present a rapid and accurate method of calibrating seismic systems using a random binary calibration signal and cross-spectral techniques. The complex transfer function obtained from the cross spectrum is least-squares fit to the ratio of two polynomials in s(s = iω) whose degrees are determined by a linear systems analysis. This provides a compact representation of the system frequency response. We demonstrate its application to two seismic systems, the IDA and SRO seismomenters. This method yields calibrations to an accuracy of better than 1 per cent in amplitude and 1° in phase.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 35 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |