doi: 10.1364/ao.26.005276
pmid: 20523516
The reference wavelength method is used for a two-color pyrometer and, with the reference wavelength method, an analytical formula of the ratio temperature for the two-color pyrometer is derived. For one channel of the two-color pyrometer, with a triangular spectral response of 0.1-micro m FWHM and 2.0-micro m peak wavelength, the effective wavelength and the correction factors with several reference wavelengths are determined. By fitting the curves of the effective wavelength and the correction factor to simple functional forms of temperature, the radiance errors for both cases are calculated. Also, it is found that the correction factor determined in a single-color pyrometer can be used directly in one channel of the two-color pyrometer without additional calculation.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 19 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |