pmid: 1066344
We have constructed an electronically controlled respirator from three commercially available components: a positive-pressure ventilator, a recorder pen motor, and a differential amplifier. Using negative feedback derived from a tracheal pressure signal, the instrument functions as a servo respirator which provides precise control of tracheal pressure. The system's power and response characteristics are well suited for ventilation of anesthetized cats and dogs. The servo respirator can be used as an externallycontrolled respiratory pump which provides flexibility in selection of the parameters of the ventilatory cycle. Alternatively, it can function as a “demand” respirator which generates transthoracic pressure proportional to efferent respiratory discharge.
Ventilators, Mechanical, Electronics
Ventilators, Mechanical, Electronics
| citations This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 39 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |