pmid: 39088619
An ever-present limitation of transmission electron microscopy is the damage caused by high-energy electrons interacting with any sample. By reconsidering the fundamentals of imaging, we demonstrate an event-responsive approach to electron microscopy that delivers more information about the sample for a given beam current. Measuring the time to achieve an electron count threshold rather than waiting a predefined constant time improves the information obtained per electron. The microscope was made to respond to these events by blanking the beam, thus reducing the overall dose required. This approach automatically apportions dose to achieve a given signal-to-noise ratio in each pixel, eliminating excess dose that is associated with diminishing returns of information. We demonstrate the wide applicability of our approach to beam-sensitive materials by imaging biological tissue and zeolite.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 16 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |