doi: 10.1118/1.596444
pmid: 2280742
Sources of variations in the measurement of half‐value layer (HVL) in film–screen mammography are evaluated. Two principal sources are the energy responses of ionization chambers and the elemental compositions of the 1100 aluminum‐alloy attenuation materials. Methods in measuring the thickness of attenuation materials are also an important potential source of error. These factors as well as the influence of geometry of measurement, position of measurement in the field, and precision of radiation output are examined. Recommendations for consistency in measurement are provided.
Data Interpretation, Statistical, Biophysics, Humans, Female, Biophysical Phenomena, Mammography
Data Interpretation, Statistical, Biophysics, Humans, Female, Biophysical Phenomena, Mammography
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 42 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |