doi: 10.1111/pace.13868
pmid: 31891195
AbstractA recently published case documented left ventricular (LV) inhibition of a Boston Scientific device by a premature complex (VPC) that was undetected by the right ventricular channel. We have observed a similar response in two patients with a BIOTRONIK cardiac resynchronization device also designed with LV sensing. A spurious response simulating that of the two true cases was also observed in a third patient with a defective LV lead which created isolated false signals. The responses of the BIOTRONIK devices were identical to that of the previously reported findings with the Boston Scientific device. The observations provide insight into the timing function of cardiac resynchronization devices designed with LV sensing.
Electrocardiography, Humans, Equipment Failure, Cardiac Resynchronization Therapy Devices, Ventricular Premature Complexes
Electrocardiography, Humans, Equipment Failure, Cardiac Resynchronization Therapy Devices, Ventricular Premature Complexes
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |