AbstractA 52‐year‐old man experienced a subcutaneous implantable cardioverter‐defibrillator (S‐ICD) inappropriate shock due to electrode tip decubitus. The device, implanted two years before with a three‐incision technique, was extracted, and a new electrode was implanted along the contralateral parasternal line with a two‐incision technique, in a one‐stage procedure. One‐year follow‐up was eventless. Early S‐ICD electrode extraction and reimplantation during the same procedure is effective and should be considered as soon as initial signs of decubitus appear to avoid inappropriate shocks. A two‐incision technique should be preferred to reduce the risk of electrode tip decubitus.
Male, Subcutaneous Tissue, Treatment Outcome, Electric Countershock, Humans, Middle Aged, Defibrillators, Implantable
Male, Subcutaneous Tissue, Treatment Outcome, Electric Countershock, Humans, Middle Aged, Defibrillators, Implantable
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 13 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |