Powered by OpenAIRE graph
Found an issue? Give us feedback
image/svg+xml Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Closed Access logo, derived from PLoS Open Access logo. This version with transparent background. http://commons.wikimedia.org/wiki/File:Closed_Access_logo_transparent.svg Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Artificial Organsarrow_drop_down
image/svg+xml Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Closed Access logo, derived from PLoS Open Access logo. This version with transparent background. http://commons.wikimedia.org/wiki/File:Closed_Access_logo_transparent.svg Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao
Artificial Organs
Article . 2018 . Peer-reviewed
License: Wiley Online Library User Agreement
Data sources: Crossref
versions View all 2 versions
addClaim

Evaluation and Comparison of Hemodynamic Performance of Three ECLS Systems in a Simulated Adult Cardiogenic Shock Model

Authors: Akif Ündar; Shigang Wang; Morgan Moroi; Allen R. Kunselman; Christoph E. Brehm;

Evaluation and Comparison of Hemodynamic Performance of Three ECLS Systems in a Simulated Adult Cardiogenic Shock Model

Abstract

AbstractThe objective of this study was to evaluate three commercially available ECLS systems with rotary pumps in terms of circuit pressure, pressure drop, perfusion modes, and hemodynamic energy transmission in a simulated adult cardiogenic shock model. One circuit consisted of a Cardiohelp system, which included a Cardiohelp console and HLS Module Advanced 7.0 tubing set with integrated centrifugal pump and oxygenator. The alternative circuit was composed of a Quadrox‐D Adult oxygenator connected in series with either an i‐cor diagonal pump and console or a Rotaflow centrifugal pump and console. The circuit was primed with lactated Ringer's solution and packed red blood cells (hematocrit 40%). The trials were conducted at flow rates of 1‐5 L/min with pseudo patient pressures of 60 mm Hg and 80 mm Hg. Pulsatile flow was tested when using the i‐cor system. Mean pre‐oxygenator pressure and pressure drop across ECLS circuit (including oxygenator and arterial tubing) were lower when using the Cardiohelp system as compared to the Rotaflow and i‐cor systems (P < 0.01). The i‐cor system was able to deliver more hemodynamic energy to the pseudo patient because of its ability to produce pulsatile flow (P < 0.01). The Cardiohelp HLS Module Advanced 7.0 integrated oxygenator had a lower resistance than the Quadrox‐D oxygenator. Although the compact Cardiohelp system had a better hemodynamic performance when compared to Rotaflow and i‐cor systems, the pulsatile flow of the i‐cor system delivered more hemodynamic energy to the pseudo patient. This may render more physiological benefits in high‐risk patients on ECLS.

Keywords

Time Factors, Hemodynamics, Models, Cardiovascular, Shock, Cardiogenic, Blood Pressure, Prosthesis Design, Extracorporeal Membrane Oxygenation, Pulsatile Flow, Materials Testing, Humans, Heart-Assist Devices, Oxygenators, Membrane

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    17
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Top 10%
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Top 10%
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Top 10%
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
17
Top 10%
Top 10%
Top 10%
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!