The development and testing results of a new current transformer (CT) test set at Physikalisch-Technische Bundesanstalt are described. The operation range for the ratio based bridge is intended for frequencies from 16.7 Hz up to 20 kHz. Equal setups of magnetically shielded current comparators with primary windings for rated currents from 100 mA up to 5 A convert the secondary currents of the standard CT and of the transformer under test (TUT) into proportional voltages. The errors of the TUT are calculated from the subsequent measurement of the complex voltage ratio. This current bridge and the voltage-ratio-based method allow comparing CTs with nonequal ratios. At 50 Hz, the measuring system offers an extended measurement uncertainty of less than 2 ppm or μrad, respectively. The bridge is verified for use with uncertainties of below 200 ppm or 400 μrad at a frequency of 20 kHz.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 24 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |