An interlaboratory comparison at 10 V DC has been made among 16 national, industrial, and military standards laboratories in North America. The comparison used a set of four traveling Zener dc reference standards. A pivot laboratory made measurements at the beginning, at the end, and at four other times during the comparison. The measured differences and their uncertainties are reported and used to demonstrate equivalence between each participant and the pivot, and between each participant and the National Institute of Standards and Technology (NIST). All of the differences fall within two parts in 10/sup 8/.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 8 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |