A master-slave type remote ultrasound diagnostic system was developed. This paper presents its controller. The controller has impedance control capability for the master and slave manipulators' positions. And it has also continuous path (CP) control capability for the slave manipulator's orientation. The impedance control allows the master and slave manipulators to move according to the motion control law, and allows the medical doctor to feel feedback force. Manipulability changes according to the impedance parameters. This paper makes clear the relationship of impedance parameters with manipulability and response speed, and proposes a method to determine appropriate impedance parameters. Experimental results demonstrate that the medical doctor can perform diagnosis successfully using the proposed system.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 10 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |