doi: 10.1093/rpd/ncn141
pmid: 18448439
Performance measurements of 30 mammographic installations were carried out in order to see the current level of image quality and breast doses.The half of the systems tested in this survey indicated automatic exposure control and beam collimation problems. Film processing and dark room conditions were not optimum for the majority of the installations. Image quality phantoms were exposed by the user and team of the survey at each visited center. Lower breast doses were obtained at equal image qualities for the radiographs of the team indicating the importance of adequate handling of some exposure parameters. Corrective actions were suggested to each installation and necessary guidance was advised for the implementation of routine quality control activities.
Quality Control, Radiographic Image Enhancement, Humans, Female, Breast, Radiation Dosage, Mammography
Quality Control, Radiographic Image Enhancement, Humans, Female, Breast, Radiation Dosage, Mammography
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 4 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |