doi: 10.1093/bja/46.4.298
pmid: 4615729
SUMMARY Recently introduced plastic swivel connectors (both autoclavable and disposable) were examined under laboratory conditions for gas leakage. The leak from the autoclavable connectors before autoclaving nearly always exceeded 5% Although this leak was diminished after autoclaving, the swivelling property was greatly decreased. The position of maximal leakage was found to vary with the position of the connector on the tracheal or tracheostomy tube adaptor. The magnitude of the leaks obtained varied, only the disposable connectors complying with our own criteria. Comparison of the plastic connectors was made with one type of metal swivel connector which was also found to be acceptable for clinical use.
Intubation, Intratracheal, Pressure, Sterilization, Tracheotomy, Disposable Equipment, Plastics, Polytetrafluoroethylene, Intermittent Positive-Pressure Breathing
Intubation, Intratracheal, Pressure, Sterilization, Tracheotomy, Disposable Equipment, Plastics, Polytetrafluoroethylene, Intermittent Positive-Pressure Breathing
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 1 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |