image/svg+xml Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Closed Access logo, derived from PLoS Open Access logo. This version with transparent background. http://commons.wikimedia.org/wiki/File:Closed_Access_logo_transparent.svg Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Journal of Endourolo...arrow_drop_down
image/svg+xml Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Closed Access logo, derived from PLoS Open Access logo. This version with transparent background. http://commons.wikimedia.org/wiki/File:Closed_Access_logo_transparent.svg Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao
Journal of Endourology
Article . 2009 . Peer-reviewed
License: Mary Ann Liebert TDM
Data sources: Crossref
versions View all 2 versions
addClaim

Foley Catheter Characteristics: Predicting Problems

Authors: Kari, Hendlin; Jason, Meyers; Manoj, Monga;

Foley Catheter Characteristics: Predicting Problems

Abstract

To evaluate the physical properties of urethral catheters that are predictive of ease of placement and risk of urethral trauma: Stiffness, buckling, and friction forces.Eleven Bard Lubri-Sil and 10 Kendall Dover 16F Foley catheters were tested using a linear motion stage driven by a stepper motor with a resolution of 5 microns/step or by hand. Force was measured at a sampling rate of 5 Hz with a Wagner FDIX digital force gauge. Friction force was measured by pulling each Foley catheter at a rate of 2.5 mm/s through a 15.6F hole drilled in 36-mm thick biologic material (Oscar Meyer Beef Light bologna 12 oz package). Ease of advancement was measured by the force necessary to maneuver a catheter around a 90-degree turn in a bulbar urethra modeled by 5/16 inch inner diameter silicone tubing and lubricant. Buckling forces were measured at the tip (8 mm), from the distal end of the balloon to the tip (30 mm), and along a 20 cm section of the shaft using the appropriate size force gauge.No significant difference was noted for friction force. The Kendall catheters (0.0686 +/- 0.102 lb(f)) took more force to advance around a 90-degree turn, compared with the Bard (0.602 +/- 0.110 lb(f), P = 0.003). Buckling force along the shaft and from the balloon to the tip was significantly greater for the Bard (0.051 +/- 0.02 lb(f), P < 0.001 and 0.892 +/- 0.166 lb(f), P < 0.001) compared with the Kendall (0.027 +/- 0.007 lb(f) and 0.803 +/- 0.129 lb(f)) catheters. Stiffness of the tip was significantly greater for the Kendall (2.644 +/- 0.552 lb(f), P < 0.001) compared with the Bard (2.101 +/- 0.437 lb(f)) Foley catheters.Traumatic catheterization is a major cause of urethral stricture. Catheter stiffness may increase the risk of urethral injury and subsequent stricture formation. Ease of advancement and tip stiffness may be important at predicting clinical performance of urethral catheters.

Related Organizations
Keywords

Animals, Cattle, Urinary Catheterization, Biomechanical Phenomena

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    3
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Average
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Average
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Average
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
3
Average
Average
Average
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!