doi: 10.1063/1.4906758
pmid: 25638144
Improvements to equipment lifetime and measurement reproducibility have been made by modifying the actuating mechanism of a diaphragmless shock tube that is used for high temperature gas kinetic studies. The modifications have two major benefits while retaining the simplicity of the original apparatus. First, the reproducibility of shock wave generation has been greatly improved and is demonstrated with 50 nearly identical experiments on the dissociation of cyclohexene at T2 = 1765 ± 13 K and P2 = 120 ± 1 Torr, demonstrating the capability for signal averaging over many experiments. Second, the lifetime of the bellows which forms the heart of the actuator is considerably improved, significantly increasing the time between replacements.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 21 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |