doi: 10.1063/1.2392722
The authors demonstrate detection and mapping of brain magnetic fields evoked by auditory stimulation with a noncryogenic magnetometer based on spin precession of potassium atoms in spin-exchange-relaxation-free regime. Optical readout using a photodiode array allows flexibility in detector placement while using common elements for most components of the multichannel system. Absence of a cryogenic dewar eliminates magnetic Johnson noise from radiation shields and allows the use of a compact magnetic shield with a high shielding factor. The magnetometer sensitivity is currently equal to 3.5fT∕Hz1∕2 at 10Hz, similar to superconducting quantum interference device magnetoencephalography systems.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 383 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 0.1% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 1% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |