doi: 10.1063/1.1291186
The SNXM [1] was first installed at the European Synchrotron Research Facility in November 1998. It has been designed for water-window operation at a spatial resolution of about 10 nm and in its final form will comprise a Zone Plate focusing X-ray onto a cylindrical collimator 10–20 nm in diameter, made by drilling an AFM tip, with its exist aperture within a few nm of the specimen surface. The operation of the microscope may be loosely defined to be in near-field in analogy with the optical SNOM e.g. [2]. Point to point resolution equal to the collimator diameter is expected for specimens up to 200 nm thick. The collimator to surface separation is also monitored by the AFM scanning tip. Simultaneous signals are available from X-ray transmission and surface topography. A progress report is given limited by the current availability of high energy X-rays (3–6 Kev).
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |