doi: 10.1039/b718187f
pmid: 18575644
A new procedure is presented for the measurement of plasminogen activators using a thickness shear mode sensor and a modified version of the fibrin plate assay at the micro-scale. Separate, well-mixed solutions of the substrates fibrinogen and plasminogen, and enzymes thrombin and the plasminogen activator sample were mixed together and placed on the sensor surface. The temperature and evaporation were controlled during the assay. The clot dissolution time correlated well with the quantity of the plasminogen activator in the sample. The average relative standard deviation was 12.5%.
Plasminogen Activators, Electrochemistry, Thrombin, Fibrinogen, Humans, Plasminogen, Acoustics, Blood Coagulation
Plasminogen Activators, Electrochemistry, Thrombin, Fibrinogen, Humans, Plasminogen, Acoustics, Blood Coagulation
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 4 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |