Abstract Individual monitoring services for external radiation were tested in the Czech Republic. The results of the tests show that they are dosimetry systems authorized and regularly tested in the Czech Republic having outliers for low energy region and angle of 50 degree and greater. With the highest probability, it can be stated that the processing equipment does not determine the angle of exposure reliably. These cases may have caused overestimations of Hp(10) in the special conditions occurring e.g. in interventional radiology and cardiology procedures.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 0 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |