handle: 11577/2801946
Abstract The pulse shape discrimination (PSD) between neutrons and gamma rays in liquid scintillators is studied by using the charge integration method with fast digitizers having different technical characteristics. The use of the Figure of Merit (FoM) to verify the PSD capability is discussed. The dependence of the FoM on the digitizer sampling rate and resolution is experimentally determined. The effects due to the type of source and the irradiation geometry are also evidenced and discussed.
Liquid scintillator; Digital signal processing; Pulse shape discrimination
Liquid scintillator; Digital signal processing; Pulse shape discrimination
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 64 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Top 10% | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Top 10% | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Top 10% |