Abstract In this study, we fabricated a multi-dimensional fiber-optic radiation sensor, which consists of organic scintillators, plastic optical fibers and a water phantom with a polymethyl methacrylate structure for the ocular proton therapy dosimetry. For the purpose of sensor characterization, we measured the spread out Bragg-peak of 120 MeV proton beam using a one-dimensional sensor array, which has 30 fiber-optic radiation sensors with a 1.5 mm interval. A uniform region of spread out Bragg-peak using the one-dimensional fiber-optic radiation sensor was obtained from 20 to 25 mm depth of a phantom. In addition, the Bragg-peak of 109 MeV proton beam was measured at the depth of 11.5 mm of a phantom using a two-dimensional sensor array, which has 10×3 sensor array with a 0.5 mm interval.
| selected citations These citations are derived from selected sources. This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | 5 | |
| popularity This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network. | Average | |
| influence This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically). | Average | |
| impulse This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network. | Average |