image/svg+xml Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Closed Access logo, derived from PLoS Open Access logo. This version with transparent background. http://commons.wikimedia.org/wiki/File:Closed_Access_logo_transparent.svg Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao The Kneearrow_drop_down
image/svg+xml Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao Closed Access logo, derived from PLoS Open Access logo. This version with transparent background. http://commons.wikimedia.org/wiki/File:Closed_Access_logo_transparent.svg Jakob Voss, based on art designer at PLoS, modified by Wikipedia users Nina and Beao
The Knee
Article . 2020 . Peer-reviewed
License: Elsevier TDM
Data sources: Crossref
versions View all 2 versions
addClaim

Catastrophic failure of biconcave unicompartmental polyethylene bearings

Authors: David, Campbell; Peter, Lewis; Luke, Mooney;

Catastrophic failure of biconcave unicompartmental polyethylene bearings

Abstract

A biconcave polyethylene (PE) bearing was developed for mobile-bearing unicompartmental knee arthroplasties (UKA) to reduce PE dislocation.A modification of the BalanSys Unicompartmental knee system with a biconcave PE and convex tibial component was used in 32 prostheses in 28 patients. Clinical outcomes and five cases of PE fracture are reported and extensively analyzed ex vivo.Visual Analogue Score of pain and satisfaction, and Knee Society Scores improved for all patients. The passive range of motion was 130°. No PE bearings dislocated. Five bearings fractured with oxidation, cracking and delamination at the thinnest central region of the PE. The combination of increased stress and decreased poly thickness was associated with increased creep. PE oxidation caused embrittlement and contributed to fractures in the thin waist of the implant.The unforeseen consequence of a novel design of a UKA that resulted in a specific mechanical and tribological mode of failure is reported. We conclude the PE failed due to a biconcave design that increased stress on the implant at its region of risk combined with a decrease of the polyethylene thickness by 1 mm in the central area.

Related Organizations
Keywords

Male, Databases, Factual, Knee Joint, Biocompatible Materials, Middle Aged, Prosthesis Design, Prosthesis Failure, Polyethylene, Humans, Female, Stress, Mechanical, Arthroplasty, Replacement, Knee, Knee Prosthesis, Oxidation-Reduction, Aged

  • BIP!
    Impact byBIP!
    selected citations
    These citations are derived from selected sources.
    This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    0
    popularity
    This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
    Average
    influence
    This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
    Average
    impulse
    This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
    Average
Powered by OpenAIRE graph
Found an issue? Give us feedback
selected citations
These citations are derived from selected sources.
This is an alternative to the "Influence" indicator, which also reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Citations provided by BIP!
popularity
This indicator reflects the "current" impact/attention (the "hype") of an article in the research community at large, based on the underlying citation network.
BIP!Popularity provided by BIP!
influence
This indicator reflects the overall/total impact of an article in the research community at large, based on the underlying citation network (diachronically).
BIP!Influence provided by BIP!
impulse
This indicator reflects the initial momentum of an article directly after its publication, based on the underlying citation network.
BIP!Impulse provided by BIP!
0
Average
Average
Average
Upload OA version
Are you the author of this publication? Upload your Open Access version to Zenodo!
It’s fast and easy, just two clicks!